The Ivoclar Programat P95 is a versatile and user-friendly furnace designed for dental laboratories and clinics. Its advanced technology and intuitive interface make it an ideal choice for processing a wide range of dental materials. The furnace features:
No. The P95 has a unique heating element geometry and software logic. Using a P90 manual will lead to programming errors. Always use the P95-specific manual .
Insert the main power cable into a grounded wall outlet. Flip the primary power switch to the "ON" position. The display screen will initialize. Operating Controls and Interface Navigation Ivoclar Programat P95 Manual
If the wire melts precisely at 961°C on the display, calibration is perfect.
| Task | Interval | Manual Page Reference | | --- | --- | --- | | Clean firing chamber with ceramic brush | Daily | p. 42 | | Inspect door seal for damage | Weekly | p. 43 | | Replace vacuum pump oil | Every 6 months | p. 55 | | Replace thermocouple | Every 2 years or 1,000 cycles | p. 61 | | Calibrate temperature offset | Annually | p. 78 | The Ivoclar Programat P95 is a versatile and
Place the unit on a level, heat-resistant surface. Allow at least 10–15 cm of space around the unit for ventilation.
offer advanced touchscreens and automated settings, many pros still prefer the manual control and tactile feedback of the P95. It provides a fundamental understanding of firing cycles that every technician should master. Final Thoughts The P95 has a unique heating element geometry
Always use the supplied tongs to place objects in the firing chamber and avoid touching hot furnace parts during or immediately after operation.
Over time, silver from alloys or contaminants from embedding investments can accumulate in the insulation, causing porcelain discoloration. Run an empty "purge" or decontamination cycle at 1000°C with the vacuum engaged for 10 minutes once a week.